ZY-5019-S 织物透湿试验仪 ZY-5019-S fabric moisture permeability tester
Instrument usage:
The principle of the moisture permeable cup weighing method is to measure the transmittance of water vapor to materials such as films, sheets, and paper by weighing the weight of the moisture permeable cup.
Compliant with standards:
GB 19082-2009 Technical Requirements for Medical Disposable Protective Clothing 5.4.2 Moisture Permeability;
GB/T 12704-1991 Method for the Determination of Moisture Permeability of Fabrics - Moisture Permeability Cup Method 6.1 Method A - Moisture Absorption Method;
GB/T 12704.1-2009 Textiles - Test Method for Moisture Permeability of Fabrics - Part 1: Moisture Absorption Method;
GB/T 12704.2-2009 Textiles - Test Method for Moisture Permeability of Fabrics - Part 2: Evaporation Method;
GB 1037, GB/T16928, ASTM E96, ASTM D1653, TAPPI T464, ISO2528, DIN 53122-1, JIS Z0208, YBB 00092003.
Product features:
1. Large screen color touch screen, Chinese and English menu operation interface;
2. The standard testing environment module and moisture permeability testing module are designed in a separate way, and the internal wind speed of the box will not affect the moisture permeability testing module, and the temperature and humidity are more uniform;
3. The standard test environment module can be used separately as a constant temperature and humidity chamber;
4. The standard test environment module meets the standard requirements, and the specified temperature and humidity are reached again within 3 minutes after closing the box door;
5. The moisture permeability testing module is equipped with a dedicated air duct and a variable frequency control method to make wind speed, temperature, and humidity more stable;
6. The moisture permeability testing module is equipped with high-precision wind speed sensors to monitor the wind speed of the test air duct;
7. The equipment can be tested by either placing desiccants in a test cup for external humidity control for weight gain, or by testing the water in the cup for external humidity control for weight loss experiments
8. Intermittent weighing equipment that meets standard requirements, equipped with an automatic intermittent weighing testing mechanism, and the weight sensor can automatically reset during the testing process
9. 2-layer cup holder design, standard with national standard moisture permeable cups;
10. Rotating sample holder for more uniform sample testing results;
11. Refrigeration system: The original imported French "Taikang" refrigeration unit ensures long-term stable operation of the entire machine;
12. Utility model patented product, patent number: ZL 2020 2 2347247.4;
13. Supporting the operating system independently developed by Zhejiang Sangong Instrument Co., Ltd.: testing instrument general software V1.0.0.
Technical parameters:
1. Standard test environment module:
1.1. Temperature range: 15 ℃~50 ℃, ± 0.1 ℃;
1.2. Humidity range: 30-98% RH, ± 1% RH;
1.3. Fluctuation/uniformity: ≤ ± 0.5 ℃/± 2 ℃, ± 2.5% RH/+2-3% RH;
1.4. Control System: Controller LCD display touch temperature and humidity controller, single point and programmable control;
1.5. Time setting: 0H1M~999H59M;
1.6. Sensor: Dry wet bulb platinum resistor PT100;
1.7. Heating system: nickel chromium alloy electric heating heater;
1.8. Refrigeration system: Original imported from France "Tai