HotSearch:
Products
ZY-8820 Respirator leak test system
1. Application: It is used to determine the leakage rate of particles in masks, face masks, respirators and gas masks.
Second, meet the standards
GB 2626-2019 Respiratory protective equipment Self-priming filter anti-particulate respirator 6.4 Leakage
GB 2890-2009 Respiratory protection Self-priming filter gas mask and other standards.
3. Product features:
1. The size of the detection chamber is 2000mm long, 2000mm wide, 2500mm high, and the volume is 10m³. The silo body is made of stainless steel frame and tempered glass.
2. The stable flow net at the top of the detection chamber makes the gas form an irregular swirling flow, accelerate the gas mixing, so that no dead angle is left in the cabin, and the pollutant uniformity required by the experiment is reached after several cycles.
3. The overall structure of the inspection chamber is mainly composed of cabin, air purification system, exhaust duct and air sampling system.
4. Sampling (inlet) sample hole: 304 stainless steel material, a total of four, ball valve control, with hose.
5. Guardrail: Install guardrail on the sampling hole surface.
6. Lighting: 2 LED fluorescent tubes (lighting lamps and fans are installed on the top of the test cabin, fixed inside and outside with Teflon material, and sealed with non-releasing imported glass glue).
7. Air exhaust system (external purification system): ensure that the air discharged from the sewage outlet can meet the emission standards.
8. A set of particle concentration detection device.
4. Technical parameters:
1. The volume of the test chamber: 10m³
2. The size of the detection chamber (L×W×H): 2000mm×2000mm×2500mm, +0.5m3 deviation allowed
3. Fogging dust source: NaCl, DEHS
4. Air volume of NaCl particulate aerosol: not less than 100L/min
5. NaCl particle concentration: 4mg/m³~12mg/m³
6. Aerodynamics of NaCl particles: particle size distribution 0.02~2um; mass median diameter is about 0.6um
7. The amount of oil particles aerosol generated: not less than 100L/min
8. Oil particle concentration: 20mg/m³~30mg/m³
9. Aerodynamics of oil particles: particle size distribution 0.02~2um; mass median diameter is about 0.3um
10. Dynamic range of particle detector: 0.001~200mg/m³; accuracy 1%; response time is not more than 500ms
11. Adjustment range of sampling pump: 0.50~4L/min
12. Particle sampling flow rate: 1~2L/min
13. Power requirements: 220V, 50Hz, 2KW
14. Air source requirements: 0.5MPa, 120L/Min
15. Overall dimensions: detection chamber (L×W×H) 2600mm×2600mm×2600mm
Test host (L×W×H): 880mm×520mm×1400mm
16. Product weight: about 420Kg for testing warehouse/40Kg for testing host
  • 上一个产品:
  • 下一个产品: ZY-8816 Respirator dead cavity tester
    ZY-8816 Respirator dead cavity tester
  • Homepage   |    About   |    Advantage   |    Products   |    News   |    Support   |    Customer   |    Jobs   |    Contact   |    中文版
    Add: A109# Diexinju Huanghe Road, Huanjiang Town, Dongguan City, Guangdong Province ,China.
    Tel:+86-0769-82307158(总机) +86-769-88762868 Fax: +86-769-89768088 Email: zhuoyueyiqi@163.com
    Copyright © Dongguan Zhuoyue Equipment Co.,Ltd. & KTS Equipment technology co.,ltd All Rights Reserved.
    Web Design: CNBASHU

    Zhuoyue Wechat