product manual:
The ZY series mechanical impact testing machine is mainly suitable for impact testing of various robot controllers/impact testing of electronic products, aerospace, shipbuilding, military industry, auto parts, transportation and other fields. By choosing different waveform generators, impact tests such as half sine wave, trailing peak sawtooth wave or square wave can be carried out.
The equipment meets: GB/T2423-2008, GJB150, GJB360, GJB548, GJB1217, MIL-STD-202F, IEC-68-2-27, MIL-STD-810F, MIL-STD-883E and other test specifications.
Main performance:
1. A free-fall impact machine with hydraulic lifting and high-pressure oil braking,
2. Adopt gas-liquid pressurization and strong friction brake to prevent secondary impact braking;
3. Impact measuring instrument: single-channel data acquisition, which can perform acceleration data measurement, storage, report printing, data callback, database management and other functions of two channels at the same time.
4. The rubber module is used to generate a wide range of half-sine wave pulses with any action time
5. The equivalent drop test of the packing box can be done.
Specifications:
Main parameters of ZY-1033-T50 impact machine
Work platform:500×600 mm
Maximum load:50KG
Peak acceleration:Half sine wave: 20-- 600 G
Back peak sawtooth wave: 30-100G
Square wave: 30---100G
Pulse duration
Half sine: 1---30ms
Back peak sawtooth wave: 6-18ms
Square wave: 6--18ms
Equipment size:1200×1050×2450mm
Total equipmen:2000 kg
measuring system
Input channel:2 channels
Sampling frequency:192KHz
voltage range:-10V--+10V
Coupling method:AC/DC/ICP
Pulse duration:0.5-100ms
Maximum acceleration:5000G
Communication Interface:USB2.0
Support standard:ISO, MIL-STD-810, user-defined standard
operating system:Microsoft Windows2000/XP
System acceleration sensor:
Brand:B&W
model:22100
output method
Charge type:
Sensitivity:3.93pC/g
Frequency Range:0.5---12KHz
Acceleration range:±2500G
working environment:-40~+160℃
other instructions:power supply
AC 3-phase 380V±10% 5KVA
Compressed air:0.5-0.8Mpa
temperature:RT~40℃
humidity:25℃<85RH%
reliability
The average failure rate of equipment is ≥5000 hours;
Half sine wave rubber is used continuously ≥10000 times, stored for five years under normal temperature environment
Applicable standards:GB2423.5, GJB150.18, GJB360.23, IEC68-2-27,MIL-STD-202F, MIL-STD-810B, etc.
Verification standard:JJG541-2005